Engineered for versatility, the compact E100m provides advanced technology for all your ventilation needs.
Back to main E100m page
View Interactive Guide for the E100m
   
The Newport E100M offers built-in valve and versatility in a simple to use ventilator with a history of exceptional reliability.

Innovative new features such as Advanced Trigger Control and Auto-Set Alarms make ventilator management easier and can improve patient care. In addition to comprehensive alarms and monitoring, the E100M offers:
Volume or Pressure Ventilation
Back-up Ventilation
DuoFlow PLUS (with Time-Limited Demand Flow)
8 hour Internal Battery Back-up
Built-in Air/Oxygen Mixer

The compact size and advanced features of the E100M allow it to easily fit in those areas of the hospital where space is limited and ease of use is appreciated; for emergency room, intra-hospital transport, post anesthesia recovery, etc. For subacute and long term care applications the E100M offers improved comfort for patients and low cost of ownership for facilities.

Our pioneering, interactive training aid for the E100M will simplify training and speed equipment proficiency. Clinical and Technical support is available 24/7 - peace of mind is just a phone call away! Contact us today and to see how simple the E100M is- to own and use.

   



Modes & Settings

 
  Pressure or Volume: A/CMV
  SMIV
  SPONT + APNEA A/CMV
  SPONT
  FlO2: .21-1.0
   (Mandatory Flow): 1 - 100 L/min
   (Inspiratory Flow): 0.1 - 3.0 sec
  VT Tidal Volume: 5 - 5000 ml
  ƒ (Frequency Rate): 1 - 120 b/min
  PEEP/CPAP: 0 - 25 cmH2O
  P limit (Pressure Limit): 0 - 120 cmH2O
  P trig (sensitivity): -10 - 25 cmH2O
  Automatic Trigger Control: On/Off
  Time Limited Demand Flow: 0 - 1.0 sec
  Paw Alarm: 3 - 118 cmH2O
  Paw Alarm: 5 - 120 cmH2O
  Auto-Set Alarm: On/Off
  Apnea: 0,15,30 seconds
  NEB Gas Source: On/Off
  PROX/PATIENT Outlet
Monitoring Selection:
On/Off
  Manual Inflation: Operator Controlled 4 sec Max.
  SPONT 2: Adjustable 0 to 15+L/min
  SPONT 1: 0 L/min or 8 L/min
  Pressure Guage: -10 to +120 cmH2O
     
  Monitored Parameters/Indicators
  Tidal Volume: Paw  Peak
  I:E Ratio:          Mean
  ƒ total:          Base
  Battery in Use: tl tE (Max 4:1)
  Battery Charge %: Sppont Detect
      
  Alarms
 
  Audible and Visual: Paw
  Paw
  Battery
  Low Baseline ( Paw)
  High Baseline
( Paw / Paw)
     
  Audible Only: Fault
  Gas Supply source failure
  Power Failure
  Alarm Silence Reset: 60 Sec
 
Electrical
  Power Requirments: 100, 120, VAC 0.14A Max.
  220, 240 VAC 0.07A. Max
  50/60 Hz
     
Internal Battery: 6-8 Hrs typical
(min 4 Hrs @ Max Load)
     
  Dimensions
 
  Width:

9.5 In (24.1 cm)

  Depth: 6.5in (16.5)
  Height: 10.5in (26.7 cm)
  Weight: 18lbs (8.2 kg)
     
  Agency Requirements

 
  Metts requirements of CSA 22.2 IEC 601-1-2 EMC MDD Directive 93/42/EEC

*Note: Specifications subject to change without notice.
Copyright © 2006 Newport Medical Instruments. All rights reserved.
1620 Sunflower Ave Costa Mesa, CA 92626, USA  Tel: +1 (714) 427-5811   Toll Free: 1 (800) 451-3111   Fax: +1 (714) 427-0489

   
Return to our Homepage