HT50

How do I calibrate the exhalation valve?

EZ Cal: 
- Enter Settings condition 
- Occlude the patient connection of the breathing circuit 
- Press Manual Inflation twiceUser Set Up 
- Enter User Set Up (press Silence/Reset during power up) 
- Press Int Battery button to scroll to Exhalation Valve Calibration 
- Occlude the patient connection of the breathing circuit 
- Press 5 to start the calibration

 