How do I calibrate the exhalation valve?
EZ Cal: - Enter Settings condition - Occlude the patient connection of the breathing circuit - Press Manual Inflation twiceUser Set Up - Enter User Set Up (press Silence/Reset during power up) - Press Int Battery button to scroll to Exhalation Valve Calibration - Occlude the patient connection of the breathing circuit - Press 5 to start the calibration
|