HT50

Does the HT50 provide continuous flow through the circuit for the pediatric patient to breathe from spontaneously if the patient fails to trigger the ventilator?

The HT50 Ventilator delivers 7.5 L/min of continuous flow through the circuit only when PEEP is set >0. When PEEP is in use, the 7.5 L/min flow helps to keep the baseline (PEEP) pressure in the breathing circuit stable during exhalation so that the trigger may be set very sensitively without auto-triggering. A sensitive trigger setting makes it easier for the patient to trigger the ventilator with very small efforts. (This flow is not needed to stabilize PEEP and minimize auto-triggering at very sensitive trigger settings when PEEP = 0 [no PEEP].) HT50 trigger may be set as sensitive as 0.1 cmH2O, can be adjusted in 0.1 cmH2O increments and is easy for very small patients to trigger.

 